Medistore
DYMIND DH36 Auto Hematology Analyzer

Standard Delivery

Delivery in 2-3 days

1 Year Warranty

Manufacturer warranty

Easy Returns

7 days return policy

Dymind

DYMIND DH36 Auto Hematology Analyzer

No ratings
|(0 reviews)|SKU: MS-528
৳ 360,000

Call us for order by Phone: +880 1988 121212

Product Type: Auto Hematology Analyzer
Brand DYMIND
Model DH36
Origin China
Warranty 01 Year
Weight NA

 

In Stock
Quantity:
1

DYMIND DH36 Auto Hematology Analyzer is an analyzer for the thought laboratory that prioritizes character results. Equipped gone a large, color LCD display, intrinsical thermal printer and frightful storage accomplish, our fan-satisfying hematology analyzer brings the convenience of accord and exactitude of the suggestion laboratory right into your follow, providing most worth back low twist of reading of reading prices.

 

Product feature illustration

Features

  • 3-part differentiation, 21 parameters
    Only 9ul sampling volume
    Up to 60 samples per hour
    50,000 sample results, including parameter, histogram, and patient information
    Venous whole blood, anticoagulant peripheral whole blood and pre-diluted bloodTechnical Specifications:Principles
    Electrical impedance method (WBC/RBC/PLT) and cyanide free (HGB) method
    Pulse baseline tracking technology, waveform discrimination technology, digital sheath flow technology and intelligent floating threshold technology
    Dual-channel is used for countingParameters
    WBC, Lym%, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,P-LCC , P-LCR, 3 histograms, alarm for abnormal erythrocytes, leukocyte, and plateletThroughput
    Up to 60 samples per hour
    Sample Volume Minimum volume 9 ulProduct feature illustrationCarryover Repeatability Linearity
    WBC ≤0.5% ≤2.0% (7-15×109/L) 0.00-300×109/L
    RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
    HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
    PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/LData Storage Capacity
    Up to 50,000 results including parameter, histogram, and patient informationCommunication
    Four USB ports and Bi-directional LIS connectionOperating Environment
    Temperature: 15oC-30oC
    Humidity: 30-85%
    Air Pressure: 70-106 kPaPower Requirement
    A.C. 100-240V ≤300VA 50/60HzDimension and Weight
    410mm(L) × 353mm(W) × 471mm(H)
    Weight: 25kg

Product feature illustration