CROMATEST hcg/Pregnancy Latex Serology (Antigen) Only Vial
Packaging Size:
Number Of Test : 100
Origin: Spain
Brand : Cromatest / Linear

Standard Delivery
Delivery in 2-3 days
1 Year Warranty
Manufacturer warranty
Easy Returns
7 days return policy
Packaging Size:
Number Of Test : 100
Origin: Spain
Brand : Cromatest / Linear
hCG-Latex Test is a rapid slide agglutination procedure, developed for the direct detection of human Chorionic Gonadotropin (hCG) in urine.
The assay is performed by testing a suspension of latex particles coated with monoclonal anti-hCG antibodies against unknown samples1. The presence or absence of a visible agglutination, indicates the presence or absence of hCG in the sample tested.

hCG-Latex Reagent. Suspension of polystyrene latex particles coated with monoclonal anti-hCG antibodies in a buffered saline solution. Contains 0.95 g/L of sodium azide.
Human urine with a hCG concentration ≥ 1600 IU/L. Contains
0.95 g/L of sodium azide.
Animal serum. Contains 0.95 g/L of sodium azide.
hCG-Latex Reagent. Suspension of polystyrene latex particles coated with monoclonal anti-hCG antibodies in a buffered saline solution. Contains 0.95 g/L of sodium azide.
Human urine with a hCG concentration ≥ 1600 IU/L. Contains
0.95 g/L of sodium azide.
Animal serum. Contains 0.95 g/L of sodium azide.
Precautions: Components of different human origin have been tested and found to be negative for the presence of antibodies anti-HIV 1+2 and anti- HCV, as well as for HBsAg. However, the controls should be handled cautiously as potentially infectious.
Warning: The reagents in this kit contain sodium azide. Do not allow to contact with skin or mucous membranes.
2610005, kit 50 tests.
hCG-Latex Reagent, 1×0.5 mL Positive control, 1×0.5 mL Negative control, 3 Test cards and 1×50 disposable pipettes.
2610015, kit 100 tests.
hCG-Latex Reagent, 1×0.5 mL Positive control, 1×0.5 mL Negative control, 3 Test cards and 2×50 disposable pipettes.
Store at 2-8ºC. Do not freeze. Frozen reagents could change the functionality of the test.
Reagent and Controls are stable until the expiry date stated on the label.
Reagent and Controls are ready to use.
Urine collected in plastic or glass containers at any time of the day. The use of the first morning urine is recommended as it generally contains the highest concentration of hormone. Samples with turbidity should be centrifuged before testing.
The samples may be stored at 2-8ºC up to two days or for longer periods at –20ºC.
Human Chorionic Gonadotropin is a glycoprotein hormone produced by the placental tissue of pregnant women, being present in serum and urine. In normal pregnancy, hCG may be detected at 7-10 days after implantation of the developing embryo, reaching maximum levels of 120.000 IU/L at 10-12 weeks of the gestation period. It is for this reason that, hCG has been considered an excellent indicator for the early detection of pregnancy.Clinical diagnosis should not be made on findings of a single test result, but should integrate both clinical and laboratory data.
The minimum detectable unit (analytical sensitivity) is of approximately 200 UI/L, referred to the 3rd International Standard of hCG 75/537 from NIBS (U.K.).
Diagnostic specificity : 99.0%
Prozone effect: No prozone effect was detected up to 250.000 IU/L.
Results obtained with this reagent did not show significant differences when compared with reference reagents.
The hemoglobin (<20 g/L) and the bilirubin (<2 mg/dL), do not interfere. Luteinizing Hormone (LH) does not interfere at concentrations £ 4 IU/L, being this concentration 20-30 fold more elevated than the maximum LH concentration found in the urine of women in the menopause period. Other substances may interfere6.
Urine from patients with trophoblastic disease such as choriocarcinoma or hydatiform mole could cause positive results7.
_ High hCG concentrations may result in either week or negative
results (prozone effect).
Qualitative Test
Reading
Nonreactive: Smooth suspension with no visible agglutination, as shown by negative control.
Reactive: Any degree of agglutination visible macroscopically (Note 4).
Positive and negative controls should be run daily following the steps outlined in the Qualitative Test, in order to check the optimal reactivity of the reagent.
The positive control should produce clear agglutination. If the expected result is not obtained, do not use the kit.
Urine may contain 50 – 5000 IU/L between 1 and 2.5 weeks of gestational age.
Negative results are expected in healthy non-pregnant women and healthy men.
In normal pregnancies a positive reaction is therefore possible 2-4 days after a missed period. If the first test should prove negative, it should be repeated a few days later, unless menstruation occurs
SOURCES OF ERROR
_ Bacterial contamination of controls and specimens as well as freezing and thawing of the hCG-Latex Reagent may lead to false positive results.
Traces of detergent in the test cards may give false positive results. Wash used cards first under tap water until reactants are removed and then with distilled water. Allow to dry in air, avoiding the use of organic solvents as they may impair the special finish on the slide.
The hCG-Latex Reagent must not be used beyond its expiry date because a prolonged storage can affect the sensitivity of the suspension.
M.E. Am. J. Obstet. Gynecol. 126(6): 678 (1976).