Cromatest Amylase Biochemistry Reagent 100 Test

৳ 10,530৳ 11,500 (-8%)

In stock

Cromatest Amylase Biochemistry Reagent

Packaging Size: 5x20 ml

Number of Test:100

Origin: Spain

Brand : Cromatest / Linear

Packaging Type: Bottle

Method: Colorimetric enzymatic method

Frequently Bought Together

Cromatest Amylase Biochemistry Reagent 100 Test

PRINCIPLE

In this direct method1,2 a-amylase catalyzes the hydrolysis of 2- chloro-p-nitrophenyl-a-D-maltotrioside (CNP-G3)  substrate at pH

6.0 forming 2-chloro-p-nitrophenol (CNP) and free glycosides.

The reaction is monitored kinetically at 405 nm by the rate of formation of the colored CNP produced, proportional to the activity of the a-amylase in the sample.

REFERENCE VALUES3

Serum And Plasma: < 86 U/L (1.43 mkat/L

Urine: < 470 U/L (7.83 mkat/L)

It is recommended that each laboratory establishes its own reference range

Best Quality Diagnostic Products Buy from Medistore

alt-gpt-br-cromatest-bio-chemistry-reagent

REAGENT COMPOSITION:

Monoreagent. MES buffer 50 mmol/L pH 6.0, calcium acetate 5 mmol/L, sodium chloride 300 mmol/L, sodium

thiocyanate 450 mmol/L, CNP-G3 2.25 mmol/L.

STORAGE AND STABILITY:

Store at 2-8 ºC.

The Monoreagent is stable until the expiry date stated on the label. Avoid contamination and recap the vials immediately after use (See Notes).

REAGENT PREPARATION

The Monoreagent is ready-to-use.

SAMPLES

Serum, heparinized plasma and urine.

Serum and plasma a-amylase is stable for 30 days at 2-8ºC. Random urine samples should be clear and precipitate free for testing. Check the pH. Urines with a pH < 5 may reduce the enzyme stability. Stable for 10 days at 2-8ºC.

Buy Hematology Analyzer Reagent at best price from Medistore

QUALITY CONTROL

To ensure adequate quality control (QC), each run should include a set of controls (normal and abnormal) with assayed values handled as unknowns.

CLINICAL SIGNIFICANCE

Amylase activity tests in serum and urine are mainly used in the diagnosis of diseases of the pancreas and in the investigation of pancreatic function.

Amylase is found chiefly in the saliva and in pancreatic tissue. Normally, small amounts of amylase are present in the blood, but with various forms of pancreatic disturbance large amounts of amylase are secreted into the blood by the pancreas.

The activity of the amylase in serum may fluctuate rapidily rising acutely during an attack and subsiding to normal levels shortly afterward.

Increased levels are found associated with acute pancreatitis, pancreatic duct obstrution, intra-abdominal diseases, mumps and bacterial parotitis.

A significant amount of the serum amylase is excreted in the urine, and as a result elevation of serum activity is reflected in the rise of urinary amylase activity. Urine amylase appears to be more frecuently elevated, reaches higher levels, and persists for longer periods.4

NOTES
  • Do not pipette by mouth, use gloves and avoid contact with skin. They both contain amylase.

ANALYTICAL PERFORMANCE

  • Linearity: Up to 1000 U/L
  • Precision
    U/L Within-run Between-run
    Mean 50.3 134.1 306 50.3 129 320
    SD 0.78 2.31 2.79 1.16 2.6 3.1
    CV% 1.55 1.72 0.65 2.32 2.01 0.96
    N 10 10 10 10 10 10

Replicates: 10 for each level.        Replicates: 10 for each level Instrument: CECIL CE 2001                for 8 days.

  • Sensitivity. Using this reagent and method an DA/min of 001 read at 405 nm is equivalent to 5 U/L of amylase activity.
  • Correlation. This assay (y) was compared with a similar commercial method (x). The results were:

N = 30      r = 0.998     y = 0.9241x + 5.237

INTERFERENCES

1.Common anticoagulants, as citrate, oxalate or EDTA inhibit the enzyme

2.Bilirubin (> 20 mg/dL) and lipids (triglycerides > 10 g/L) do not interfere.

3. Other drugs and substances may interfere3.

Total Protein Cromatest Bio Chemistry Reagent

MATERIALS REQUIRED

  • Photometer or spectrophotometer with a thermostatted cell compartment set at 37ºC, capable of reading at 405 nm.Stopwatch, strip-chart recorder or printer. Cuvettes with 1-cm pathlength.Pipettes to measure reagent and samples.

PROCEDURE

1.Preincubate working reagent, samples and controls to reaction temperature.

2.Set the photometer to 0 absorbance with distilled water

3.Pipette into a cuvette:

Reaction temperature 37ºC
Monoreagent Serum/plasma Urine 1.0 mL 20 mL

1.0 mL

10 mL

Mix gently by inversion. Insert cuvette into the cell holder and start stopwatch.

Incubate for 1 minute and record initial absorbance reading. Repeat the absorbance readings exactly after 1, 2 and 3 minutes.

Calculate the difference between absorbances.

Calculate the mean of the results to obtain the average change

in absorbance per minute (DA/min).

Calculations: Samples with DA/min exceeding 0.500 at 405 nm should be diluted 1:10 with saline and assayed again. Multiply the results by 10.

If results are to be expressed as SI units apply: U/L x 0.01667 = mkat/L

REFERENCES

  1. Winn-Deen, E.S., David, H., Sigler, G, and Chavez, R. Clin. Chem. 34 : 2005 (1988).International Federation of Clinical Chemistry (IFCC). Clin. Chem. Lab. Med. 36 : 185 (1998).Young, D.S. Effects of Drugs on Clinical Laboratory Tests. 4 thEdition. AACC Press (1995).Tiete. Textbook of Clinical Chemist, 2  Edition. Bruits CA, Ashwood ER. WB Saunders Co., 1994.

Brand

Chromatest

Linear

Be the first to review “Cromatest Amylase Biochemistry Reagent 100 Test”

Reviews

There are no reviews yet.

Main Menu

Cromatest Amylase Biochemistry Reagent 100 Test

৳ 10,530৳ 11,500 (-8%)